Disposable pressure transducer is for the continuous measurement of physiological pressure and determination of other important haemodynamic parameters. Hisern’s DPT can provide accurate and reliable blood pressure measurements of arterial and venous during cardiac intervention operations.
Indicated for pressure monitoring applications like: ● Arterial blood pressure (ABP) ● Central venous pressure (CVP) ● Intra cranial pressure (ICP) ● Intra abdominal pressure (IAP)| ITEMS | MIN | TYP | MAX | UNITS | NOTES | |
| Electrical | Operating Pressure Range | -50 | 300 | mmHg | ||
| Over Pressure | 125 | psi | ||||
| Zero Pressure Offset | -20 | 20 | mmHg | |||
| Input Impedance | 1200 | 3200 | ||||
| Output Impedance | 285 | 315 | ||||
| Output Symmetry | 0.95 | 1.05 | Ratio | 3 | ||
| Supply Voltage | 2 | 6 | 10 | Vdc or Vac rms | ||
| Risk Current (@ 120 Vac rms, 60Hz) | 2 | uA | ||||
| Sensitivity | 4.95 | 5.00 | 5.05 | uU/V/mmHg | ||
| Performance | Calibration | 97.5 | 100 | 102.5 | mmHg | 1 |
| Linearity and Hysteresis (-30 to 100 mmHg) | -1 | 1 | mmHg | 2 | ||
| Linearity and Hysteresis (100 to 200 mmHg) | -1 | 1 | % Output | 2 | ||
| Linearity and Hysteresis (200 to 300 mmHg) | -1.5 | 1.5 | % Output | 2 | ||
| Frequency Response | 1200 | Hz | ||||
| Offset Drift | 2 | mmHg | 4 | |||
| Thermal Span Shift | -0.1 | 0.1 | %/°C | 5 | ||
| Thermal Offset Shift | -0.3 | 0.3 | mmHg/°C | 5 | ||
| Phase Shift (@ 5KHz) | 5 | Degrees | ||||
| Defibrillator withstand (400 joules) | 5 | Discharges | 6 | |||
| Light Sensitivity (3000 Foot Candle) | 1 | mmHg | ||||
| Enviromental | Sterilization (ETO) | 3 | Cycles | 7 | ||
| Operating Temperature | 10 | 40 | °C | |||
| Storage Temperature | -25 | +70 | °C | |||
| Operating Product Life | 168 | Hours | ||||
| Shelf Life | 5 | Years | ||||
| Dielectric Breakdown | 10,000 | Vdc | ||||
| Humidity (External) | 10-90% (non-condensing) | |||||
| Media Interface | Dielectric Gel | |||||
| Warm-Up Time | 5 | Seconds | ||||